MA Zhi-feng, LI Xi-ze, ZHANG Feng, CHEN Cheng-ying. A Successive Approximation ADC Design for Implantable Intraocular Pressure Detection Application[J]. Microelectronics & Computer, 2017, 34(11): 135-139.
Citation: MA Zhi-feng, LI Xi-ze, ZHANG Feng, CHEN Cheng-ying. A Successive Approximation ADC Design for Implantable Intraocular Pressure Detection Application[J]. Microelectronics & Computer, 2017, 34(11): 135-139.

A Successive Approximation ADC Design for Implantable Intraocular Pressure Detection Application

  • In order to ensure the long-term stable working state and detection resolution of implantable intraocular pressure sensor, a 10 bit, 20 KS/s low power successive approximation analog to digital converter (SAR ADC) for implantable intraocular pressure detection sensor application is proposed in this paper. The DAC adopts a full capacitance array design. The symmetrical capacitor array structure improves the matching accuracy of the scaling capacitor; Comparator with three-stage pre-amplifier and dynamic latch structure ensures the offset calibration function. The circuit is implemented with SMIC 0.18 μm CMOS technology and the post-simulation results show that the effective number of bits (ENOB) can reach 9.4 bit at 20 KS/s sampling rate, and the average power consumption is only 44.7 μW.
  • loading

Catalog

    Turn off MathJax
    Article Contents

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return